 Embedded Power for Business-Critical Continuity
NFS40 Series
SIngle and triple output
Rev.02.15.07 NFS40 Medical 1 of 4
Total Power: 40 - 50W Input Voltage: 85 - 264VAC 120 - 370VDC # of Outputs: Single, triple
Special Features
* 5.0 x 3.0 x 1.2 inch package (1U applications) * Industry standard package * Overvoltage and short circuit protection * 40 W with free air convection * EN55022, EN55011 conducted noise level A * UL, VDE and CSA safety approvals * Available RoHS compliant * 2 year warranty
Safety
The NFS40-79XXJ models are approved to UL2601, CSA22.2 No. 125 and IEC601/VDE0750 standards. The NFS40-79XXJ series is for use in ordinary, patient-connect applications under the UL2601 and CSA C22.2 No. 125 standards, and is authorised for use in non-critical, non-patient-connect applications under the IEC601 standards.
The NFS40 medical series is a 40 W universal input ac-dc power supply on a 5 x 3 inch card with a maximum component height of 1.2 inches for use in medical applications. The NFS40 medical series has the same generic feature set as the standard NFS40 series but has been designed with lower safety ground leakage and higher isolation as required for medical safety approval. The NFS40 provides 40 W of output power with free air convection cooling which can be boosted to 50 W with 20 CFM of air. Standard features include overvoltage and short-circuit protection. The series, with full medical safety approval to EN60601 and UL2601 and the CE mark, meets conducted emissions EN55022 level A. The NFS40 medical series is designed for use in low power medical, dental and laboratory applications such as dialysis machines, monitoring equipment, instrumentation and infusion pump controls.
Embedded Power for Business-Critical Continuity
Specifications
All specifications are typical at nominal input, full load at 25C unless otherwise stated
Rev.02.15.07 NFS40 Medical 2 of 4
OUTPUT SPECIFICATIONS Voltage adjustability Line regulation Overshoot/undershoot Transient response Temperature coefficient Overvoltage protection Minimum output current Output power limit Short circuit protection INPUT SPECIFICATIONS Input voltage range Input frequency range Input surge current Safety ground leakage current Cold start 110 Vac, 60 Hz Cold start 230 Vac, 50 Hz 110 Vac, 60 Hz 230 Vac, 50 Hz 85-264 Vac 120-370 Vdc 47-440 Hz 10 A 22 A 18 A max. 28 A max. +5 V output on triples Vout on singles LL to HL, FL main output LL to HL, FL auxiliary output At turn-on +5 V (1.5 A to 3 A step) All outputs + 5V output (See Note 10) Primary power limited Single Multiple 5.0% 5.0% 0.2% 1.0% 0% 120 mV max dev. 500 s recovery 0.02%/C 6.25 0.75 Vout 0A 90 W input power limit Continuous Short term
EMC CHARACTERISTICS Conducted emissions Radiated emissions ESD air ESD contact Surge Fast transients Radiated immunity Conducted immunity GENERAL SPECIFICATIONS Hold-up time Efficiency Isolation voltage Switching frequency Approvals and standards (See Note 12) Weight MTBF MIL-HDBK-217E 110 Vac 230 Vac 110 Vac, 230 Vac Input/output Input/chassis 18 ms 132 ms 70% typical 4000 Vac 1500 Vac 20-10 kHz VDE0750, IEC601 EN60601-1, UL2601 CSA C22.2 No. 125 270 g (9.6 oz) 170,000 hours EN55022, FCC part 15 EN55022, FCC part 15 EN61000-4-2, level 3 EN61000-4-2, level 4 EN61000-4-5, level 3 EN61000-4-4, level 3 EN61000-4-3, level 3 EN61000-4-6, level 3 Level A Level A Perf. criteria 1 Perf. criteria 1 Perf. criteria 1 Perf. criteria 1 Perf. criteria 2 Perf. criteria 1
ENVIRONMENTAL SPECIFICATIONS Thermal performance Operating, see curve 0 C to +70C Non-operating -40 C to +85C 0 C to 50 C ambient temp., 40 W convection cooled 0 C to 50 C ambient, 50 W 20 CFM Forced air 50 C to 70 C ambient Derate linearly to 50% load Peak (30 seconds) 60 W Non-condensing Operating Non-operating 5-500 Hz 5% to 95% RH 10,000 feet max. 40,000 feet max. 0.75 G peak
Relative humidity Altitude Vibration (See Note 11)
Embedded Power for Business-Critical Continuity
Specifications Contd.
OUTPUT VOLTAGE +5.1 V (VA) +12 V (VB) -12 V
(6)
Rev.02.15.07 NFS40 Medical 3 of 4
OUTPUT CURRENT CONV. MAX (1) 3A 2A 0.35 A 3A 2A 0.35 A 3.3 A 2.6 A 1.6 A 4A 0.35 A 0.35 A 20CFM MAX. (2) 5A 2A 0A 5A 2A 0.5 A 4A 3.3 A 2A 7A 1A 1A PEAK (3) 7A 3A --7A 2.5 A --5A 4A 2.5 A 5A 0.5 A 0.5 A
RIPPLE (4) PK-PK 50 mV 120 mV 120 mV 50 mV 150 mV 150 mV 120 mV 150 mV 240 mV 50 mV 120 mV 120 mV
TOTAL REGULATION (5) 2.0% 5.0% 5.0% 2.0% +10%/-3.0% 5.0% 2.0% 2.0% 2.0% 2.0% 5.0% 5.0%
MODEL NUMBERS (13,14) NFS40-7908J
+5.1 V (VA) +15 V (VB) -15 V (6) 12 V 15 V
(7)
NFS40-7910J
NFS40-7912J NFS40-7915J NFS40-7924J NFS40-7928J
(7)
24 V (7) +5.1 V +12 V -12 V Notes
1 2 3 4 5
AC mating connector
Molex 09-50-3031 or equivalent with Molex 08-50-0105 or equivalent crimp terminal.
6 7 8 9 10 11 12 13 14
Natural convection cooling, 40 W maximum. Forced air, 20 CFM at 1 atmosphere, 50 W maximum. Peak output current lasting less than 30 seconds with duty cycle less than 10%. During peak loading, outputs may go outside of total regulation limits. Peak total power must not exceed 60 W. 50 MHz bandwidth, peak-to-peak, measured differentially. Total regulation is defined as the static output regulation at 25 C, including initial tolerance, load currents within stated limits, and output voltages adjusted to their factory settings. Also, 0.25 IA / IB 5.0 to maintain stated regulation. A minimum load of 0.5 A is required on the +5 V output to obtain full current from the negative output. Single output models have floating outputs which may be referenced as either positive or negative. Derating curve is application specific for ambient temperatures >50 1/4C, for optimum reliability no part of the heatsink should exceed 120 1/4C and no semiconductor case temperature should exceed 130 1/4C. Caution: Allow a minimum of 1 second after disconnecting the power when making thermal measurements. Although the minimum output current of the NFS40-79XXJ is 0 A, a 4 W minimum load is required to achieve design MTBF. Three orthogonal axes, sweep at 1 octave/min, 5 minute dwell at four major resonances. This product is only for inclusion by professional installers within other equipment and must not be operated as a stand alone product. The `J' suffix indicates that these parts are Pb-free (RoHS 6/6) compliant. TSE RoHS 5/6 (non Pb-free) compliant versions may be available on special request, please contact your local sales representative for details. NOTICE: Some models do not support all options. Please contact your local Artesyn representative or use the on-line model number search tool at http://www.artesyn.com/powergroup/products.htm to find a suitable alternative.
DC mating connector
Molex 09-91-0600 or equivalent with Molex 08-50-0164 or equivalent crimp terminal.
PIN CONNECTIONS J1 Pin 1 Pin 2 J2 Pin 1 Pin 2 Pin 3 Pin 4 Pin 5 Pin 6 P1 Pin 1 Safety Earth Ground +12 V +5.1 V +5.1 V Return Return -12 V +15 V +5.1 V +5.1 V Return Return -15 V +Vout +Vout +Vout Return Return Return -7908J, 7928J AC Line AC Neutral -7901J AC Line AC Neutral SINGLES AC Line AC Neutral
NO. 3
Embedded Power for Business-Critical Continuity
Rev.02.15.07 NFS40 Medical 4 of 4
Mechanical Notes
A In order to meet safety requirements, a non-metallic stand-off is mandatory for one hole as specified in the mechanical drawing above. B The ground pad of the mounting hole near P1 allows system grounding through a metal stand-off. C To improve conducted noise, the ground pad of the mounting hole near the output connector should be connected with the ground pad of the mounting hole near P1. Use metal stand-offs attached to a common metal chassis. This connection also significantly attenuates common mode noise. D A standard L-bracket and cover is available for mounting which contains all screws, connectors and necessary mounting hardware. Order part number `NFS40 COVER KIT'.
Americas
5810 Van Allen Way Carlsbad, CA 92008 USA Telephone: +1 760 930 4600 Facsimile: +1 760 930 0698
Europe (UK)
Waterfront Business Park Merry Hill, Dudley West Midlands, DY5 1LX United Kingdom Telephone: +44 (0) 1384 842 211 Facsimile: +44 (0) 1384 843 355
MAXIMUM COMPONENT HEIGHT 1.2 inches
0.156 (3.96) Dia. (4 Places)
5.00 (127.00)
Asia (HK)
16th - 17th Floors, Lu Plaza 2 Wing Yip Street, Kwun Tong Kowloon, Hong Kong Telephone: +852 2176 3333 Facsimile: +852 2176 3888
NON-METALLIC STAND-OFF (SEE NOTE A) Q1
C5
J1 3.00 (76.2)
~ N ~ L F1 F 3.15A, H 250VAC P1 (SEE NOTE C)
1 T1 J2
2.550 (64.8)
For global contact, visit:
www.powerconversion.com technicalsupport@powerconversion.com
While every precaution has been taken to ensure accuracy and completeness in this literature, Emerson Network Power assumes no responsibility, and disclaims all liability for damages resulting from use of this information or for any errors or omissions.
0.22 (5.59) 4.55 (115.8)
0.22 (5.59)
ALL DIMENSIONS IN INCHES (mm)
Emerson Network Power. The global leader in enabling business-critical continuity. AC Power Connectivity DC Power Embedded Power Inbound Power Integrated Cabinet Solutions Outside Plant Precision Cooling Site Monitoring and Services EmersonNetworkPower. com
Emerson Network Power and the Emerson Network Power logo are trademarks and service marks of Emerson Electric Co. (c)2007 Emerson Electric Co.


▲Up To Search▲   

 
Price & Availability of NFS40-7908J

All Rights Reserved © IC-ON-LINE 2003 - 2022  

[Add Bookmark] [Contact Us] [Link exchange] [Privacy policy]
Mirror Sites :  [www.datasheet.hk]   [www.maxim4u.com]  [www.ic-on-line.cn] [www.ic-on-line.com] [www.ic-on-line.net] [www.alldatasheet.com.cn] [www.gdcy.com]  [www.gdcy.net]


 . . . . .
  We use cookies to deliver the best possible web experience and assist with our advertising efforts. By continuing to use this site, you consent to the use of cookies. For more information on cookies, please take a look at our Privacy Policy. X